Features and Functions 2
34980A User’s Guide 45
• To initiate a scan and store all readings in memory, press
Scan (Measure). Each time you initiate a new scan, the instrument
clears all previously stored readings. If you have not defined a scan list,
Scan (Measure) performs an internal DMM scan independent of any
channels.
• To stop a scan in progress, press and hold Scan (Measure).
To Build a Scan List From the Remote Interface
• Use the ROUTe:SCAN command to define the list of channels in the scan
list. To determine what channels are currently in the scan list, use the
ROUTe:SCAN? query command.
• To add channels to the present scan list, use the
ROUTe:SCAN:ADD
command. To remove channels from the present scan list, use the
ROUTe:SCAN:REMove command.
• To remove all channels from the scan list, send “
ROUT:SCAN (@)”.
• To initiate a scan, use the
INITiate or READ? command.
Measurements are stored in memory. Each time you initiate a new scan,
the instrument will clear the previous set of readings from memory.
• To stop a scan in progress, use the
ABORt command.