Barco RS97670/03 Projector User Manual


 
C. Calibrate Measured Values
C. CALIBRATE MEASURED VALUES
Overview
Calibrate Measured Values
C.1 Calibrate Measured Values
Necessary tools
Color meter (e.g. PhotoResearch 650 or LMT).
How to Start Calibrate Measured Values?
1. Put the color meter at the center of the image and switch to option X-Y-Z (X=Red, Y=Green, Z=Blue). (image C-1)
2. On the projector press the MENU key to activate the Menu bar.
3. Push the cursor key or to highlight Image in the menubar.
4. Push the key to pull down the Image menu.
5. Push the cursor key or to highlight Dynacolor and press ENTER to select. (image C-2)
The Dynacolor dialog box will be displayed.
6. Push the cursor key or to highlight Calibrate Mea sured Values and press ENTER to select. (image C-3)
This will start up the calibration procedure.
Projector
Screen
Color Meter
Image C-1
Put the color meter at the center of the image and switch to option X-Y-Z (X=Red, Y=Green, Z=Blue)
R5976870 BARCOREALITY SIM 5PLUS/SIM 5R 04/04/2007 229