INSTALLATION AND PREPARATION OF STARTUP
5 - 2
ETC00303(1) BINOS E e (2.0) 11/00
Rosemount Analytical
5.1 Installation Site
Be sure to observe the additional notes, safety precautions and warnings given
in the individual manuals!
The BINOS E must not operate in explosive atmosphere without supplemen-
tary protective measures !
Free flow of air into and out of the BINOS E (ventilation slits) must not be
hindered by nearby objects or walls !
The installation site for the BINOS E has to be dry and remain above freezing
point at all times. The BINOS E must be exposed neither to direct sunlight nor
to strong sources of heat.
Be sure to observe the permissible ambient temperatures (c.f. Item 20:
Technical Data). For outdoor installation, we recommend to install the BINOS
E in a protective cabinet. At least, the BINOS E has to be protected against rain
(e.g., shelter).
The BINOS E has to be installed as near as possible to the sample point, in order to avoid
low response time caused by long sample gas lines.
In order to decrease the response time, a sample gas pump with a matching high pumping rate
may be used. Eventually, the BINOS E has to be operated in the bypass mode or by an overflow
valve to prevent too high flow and too high pressure (Fig. 5-2).
Fig. 5-2: BINOS E, Bypass installation
INSTALLATION SITE
Overpressure valve
Flow meter
Exhaust
Exhaust
BINOS E
Gas sampling pump
Filter