Emerson ETC00303 Printer Accessories User Manual


 
INSTALLATION AND PREPARATION OF STARTUP
5 - 4
ETC00303(1) BINOS E e (2.0) 11/00
Rosemount Analytical
5.2.1 Pressure Sensor (Option)
It is possible to integrate up to two pressure sensors with a range of 800 - 1100 hPa.
The concentration values computed by the analyzer will then be corrected to reflect the
barometric pressure to eliminate faulty measurements due to changes in barometric pressure
(see technical data). BI NOS E front panel program will indicate whether pressure correction
is actual.
5.2.2 Gas Flow Rate / Internal Flow Sensor (Option)
The gas flow rate should be within the range 0.2 l/min to maxi. 1.5 l/min !
A constant flow rate of about 1 l/min is recommended.
The gas flow rate for BINOS E with paramagnetic oxygen sensor
is allowed to max. 1.0 l/min !
Internal flow sensor: max. 2.0 l/min !
It is possible to integrate up to three flow sensors. In this case gas flow can be shown via
BINOS E front panel program.
GAS CONDITIONING (SAMPLE HANDLING)
5.3 Gas Connections
The installed gas connections are specific to the different BINOS E. All fittings are clearly
marked. The fittings are located on the rear panel of the BINOS E instrument.
The exhaust gas lines have to be mounted in a declining, pressureless
and frost-free way and according to the valid emission legislation !
Do not interchange gas inlets and gas outlets !