Emerson ETC00303 Printer Accessories User Manual


 
INSTALLATION AND PREPARATION OF STARTUP
5 - 3
ETC00303(1) BINOS E e (2.0) 11/00
Rosemount Analytical
5.2 Gas Conditioning (Sample Handling)
The conditioning of the sample gas is of greatest importance for the successful operation of
any analyzer according to extractive method.
All gases have to be supplied to the BINOS E as conditioned gases !
When the instrument is used with corrosive gases, it is to be verified that there
are no gas components which may damage the gas path components.
The gas has to fullfil the following conditions:
It must be
free of condensable constituents
free of dust
free of aggressive constituents which are not compatible with the material of the gas
paths.
have temperatures and pressures which are within the specifications shown in “Technical
Data” of this manual.
Flammable or (potential) explosive gas mixtures may not be introduced into
the BINOS E without supplementary protective measures !
When analysing vapours, the dewpoint of the sample gas has to be at least 10 °C below the
ambient temperature in order to avoid the precipitation of condensate in the gas paths.
Suitable gas conditionning hardware may be supplied or recommended for specific analytical
problems and operating conditions.
GAS CONDITIONING (SAMPLE HANDLING)