Emerson ETC00303 Printer Accessories User Manual


 
FRONT PANEL PROGRAM
8 - 7
ETC00303(1) BINOS E e (2.0) 11/00
Rosemount Analytical
FUNCTION KEYS
During span gas adjustment, the deviation of the actual value is compared with the max.
deviation (threshold value) from the file ”BinosOEM.INI”. The menu provides the opportunity to
change the threshold value from 1 % of full scale up to the maximum threshold value.
As with zero gas adjust-
ment, you can run span
gas calibration with indi-
vidual selectable chan-
nels or with all channels
simultaneously.
Unavailable channels will
be refused.
You can set the nominal
value (set point) in the field
behind the corresponding
channel.
If the value falls below the threshold
value, the span gas calibration proce-
dure is stopped.
To enable a span gas calibration, the
maximum threshold value may be set
(if this does not work contact our Ser-
vice Support Center).