Notes:
The signal has to be stable before starting any calibration !
If you want to start the zero and the span calibration separately, or
if you want to calibrate separately each channel of an MLT analyzer/analyzer module, or
if you want to calibrate the measurement ranges separately with span gas,
you have to change to the menu "Analyzer Module Calibration" (see p. 5-15) resp. to
the menu "Calibration Procedure Setup" (p.5-9).
You can cancel any calibration procedures with the function "Cancel all running
procedures !"
Via the F2 key you will reach the menu "Analyzer Channel Status". In that menu and
its corresponding submenus you will find the status report of the current channel for
Failures
Maintenance requests
Function controls
Measurements/Alarms
Events
Hours of operation
and the operational settings of the current channel like
range settings and
response time (t
90
-time).
The menu "Analyzer Channel Status" and its submenus are displayed detailed in
section 4.1.
Menu line "Info...":
Via this menu line you can change to the submenu "State of Calibration Procedures",
where you will find the actual status of each channel of an MLT analyzer or analyzer
module.
Measure Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- State of Calibration Procedures --
Channel -1: Ready
Channel -2: Ready
Channel -4: Ready
Channel -5: Ready
Channel -3: Ready