4.1.1 Analyzer Channel Status - Status Details
Acknowledge and Clear Maintenance Requests
90003482(2) [NGA-e (MlT-Software 3.2.X)] 07/98
NGA 2000
4 - 23
Display Status... Main... Channel Calib...
TAG
F1 F2 F3 F4 F5
Failures: No
Maintenance-Requests: Yes
Temperature: 20.0 C
Operation: Ready
0.0 100.0
Range: 1
37.50 ppm CH4
0.00 50.00
Measure Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Analyzer Channel Status --
Status details...
Hours of operations: 164
Operation status: Ready
Events: No
Alarms: No
Maintenance requests: No
Function control/Service: No
Operational settings...
Failures: No
RawMeas
Measure Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Status Details --
Failures...
Function controls...
Measurements/Alarms...
Events...
Maintenance requests...
Acknowledge and clear failures !
Acknowledge and clear maintenance requests !
Acknowledge and clear function controls !
⇒ ⇒ Change to the menu
"Analyzer Channel
Status"
Press
F2
Note:
If you have solved the
reasons for the maintenance
requests reported, you
should start this function.
The menu "List of Possible
Maintenance Requests" will
be ready for new reports !
⇒ ⇒ Change to the
submenu "Status
Details"
Press
or
⇒ ⇒ Change to the line
"Acknowledge and
clear maintenance
requests !"
Press or
as often as necessary
to get the menu line
"Acknowledge and clear
maintenance requests !"
white on black.