Emerson NGA 2000 Computer Accessories User Manual


 
5 - 2
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Structure of chapter five:
At the begin of any chapter you will find the way to a certain submenu of the MLT software
starting from the line "Analyzer and I/O-module expert configuration" in the "Main Menu".
The way will be described with the software catchwords, that you have to enter one after
another to reach the corresponding submenu. At the end of the catchword listing you will
find the illustration of the LCD screen. After that you can read the set up instructions and
explanations. Sometimes the illustration will be completed by sketches or further menu
pictures.
Example: You will set up the calibration parameters for the calibration procedure of the
analyzer module.
Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Calibration parameters
Calibration procedure setup
In the menu "Calibration Procedure Setup" you can set up the parameters for the cali-
bration procedure of the zeroing and spanning ...
Further explanations and instructions will follow !
Measure Channel Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Calibration Procedure Setup --
Purge time: 10 s
Analog output during calibration: Tracking
Span ranges: Together
Valve position: Samplegas
Stability time: 30 s
Averaging time: 5 s
Procedure times-out after: 120 s