4 - 44
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Yes Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Measure Cancel ! Back... More...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Calibration Procedure Status --
Procedure status: Purging1-Wait
Current/expected gas flow: Spangas-1
Concentration in span gas units: 37.50 ppm
Last zero calibration: Success
Last span calibration: Success
Last span calibration was: Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges: 1+2+3+4
Remaining procedure time: 4 s
Last zero calibration was: Mon 06-08-1998 09:21:34
---------------------- Results ----------------------
Measure Cancel ! Back... More...
TAG
F1 F2 F3 F4 F5
43.57
-- Calibration Procedure Status --
Procedure status: Spanning-Wait
Current/expected gas flow: Spangas-1
Concentration in span gas units: 43.57 ppm
Last zero calibration: Success
Last span calibration: Success
Last span calibration was: Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges: 1+2+3+4
Remaining procedure time: 108 s
Last zero calibration was: Mon 06-08-1998 09:21:34
---------------------- Results ----------------------
⇒ ⇒ Confirm the order for
span calibration
Press
F2
to start
the spanning immediately.
Option:
Press the F4 key if you want
to cancel the order.
Notes:
• The display of this message
depends on the setup in
the expert configurations.
(see 5.1.8 p. 5-49)
• The 3 following illustrations
show the displays you can
see after starting the
spanning procedure.
⇒ ⇒ Spanning:
1. Purging-Wait
Notes:
• The procedure time de-
pends on the parameters
entered for purge time
(see 5.1.1 p. 5-9).
• The purge time must be
long enough to get a stable
signal before calibration.
• You can cancel the run-
ning calibration procedure
at any time with the F2 key.
⇒ ⇒ Spanning:
2. Spanning-Wait
Note:
The procedure time de-
pends on the parameters
entered for stability time and
averaging time.
(see 5.1.1 p.5-9)