Emerson NGA 2000 Computer Accessories User Manual


 
4.4 Calibration Procedure Status
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 37
Display Status... Main... Channel Calib...
TAG
F1 F2 F3 F4 F5
Failures: No
Maintenance-Requests: No
Temperature: 20.0 C
Operation: Ready
0.0 100.0
Range: 1
37.50 ppm CH4
0.00 50.00
Measure Status... Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas: 50.00 ppm
Span gas name: Methane
Start zero calibration procedure !
Expected zero gas: 0.00 ppm
Valves...
Valve position: Samplegas
Time & Date: 12:50:55 June 16, 1998
Measure Cancel ! Back... More...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Calibration Procedure Status --
Procedure status: Ready
Current/expected gas flow: Samplegas
Concentration in span gas units: 37.50 ppm
Last zero calibration: Success
Last span calibration: Success
Last span calibration was: Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges: 1+2+3+4
Remaining procedure time: 0 s
Last zero calibration was: Fri 05-29-1998 13:32:06
---------------------- Results ----------------------
Change to the
menu "Analyzer
Module Calibration"
Press
F5
Change to the
menu "Calibration
Procedure Status"
Press
Options:
Via the F5 key you can
change to a submenu,
where you can close all
valves or set up each valve
separately with:
zero gas or
span gas or
sample gas or
test gas.
(See sections 4.7 to 4.11,
p. 4-47...)
Change to the menu
page "Calibration
Deviations"
Press
F5