NGA-2000 System Calibration
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98 NGA 2000 Supplement - 5
It is also possible to configure the DIO-outputs via LON-Variables (i.e. SLTA-adapter or AK-
protocol).
Therefore is the LON-variable "DIOOUT_ENTRYSIG" which corresponds to DIOOUTSIGC
and evaluates enum-values of ST1NAME (enum-values 0..19), ST2NAME (enum-values
20..39) or ST3NAME (enum-values 40..59).
For digital output we have to setup variables in following order.
1. DIOSLOTIDC
2. DIOOUTNUMC
3. DIOOUTSRCC
4. DIOOUT_ENTRYSIG SYS: VALVE-1 = 20 (enum-value in ST2NAME)
SYS: VALVE-2 = 21
. .
. .
. .
SYS:VALVE-32 = 51 (enum-value in ST3NAME)
For each needed system valve we have to repeat these appropriate selections.
It is also possible to distribute the system valves onto different output modules.