NGA-2000 System Calibration
Supplement - 22 NGA 2000 90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
For example:
• Starting time for zero calibration: daily at 2:00 o´clock.
• Starting time for zero/span calibration: weekly at 10:00 o´clock.
• Starting time for program calibration: weekly at 16.00 o´clock
• Entry in menu page „times for interval operation“ are at 4/13/99 at 8.00 o´clock
• First possible calibration: 4/14/99 at 2 o´clock (then all 24 hours automatically)
We can display the calculated ACT-times in menu page „next calibration events“...
But these times will only appear when the appropriate time controlled calibration type is
enabled.
Menu1Line
Menu2Line
Menu3Line
System configuration and diagnostics...
↓
System Calibration...
↓
Times for interval operation...
↓
Next calibration events...
↓
Next Calibration Events
Zero-Cal: 2:00 14 Apr 1999
Zero/Span-Cal: 10:00 13 Apr 1999
Program-Cal: 16:00 14 Apr 1999