Emerson NGA 2000 Computer Accessories User Manual


 
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 59
Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Gasflow setup
In the menu "Gasflow Setup" you can set up the modus of gas flow through
the analyzer cell:
Parallel or
Serial
and the valve position at "Basic-Status":
Samplegas or
All closed
Notes:
The adjustment for the gas flow through the analyzer cell depends on the real
combination of the analyzer cells.
The setup is normally correct by factory setting.
You have only to change it, if you modify your configuration of cells.
The correctness of this adjustment is very important for the time controlled calibration
with valve gear (see 5.1.1 p. 5-12)
Set up parameters:
Change to the line you want with the -key or the -key.
Press the 88 -key or the -key to select the variable.
Select the parameter you want with the -key or the -key.
Confirm your adjustment with the 88 -key or
cancel and go back to the last one with the F2 key.
Measure Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Gasflow Setup --
Gasflow through analyzer cell is: Parallel
Valve position at "Basic-Status": Samplegas
Channel
Operation status: Ready
Valve position: Samplegas
5.1 Analyzer Module Setup
5.1.13 Gasflow Setup
[2nd menu page via F5 (More...)]