www.exatron.com 6-3 Chapter 6 Display Dictionary
BAD RESPONSE
The handler is using the E
XATRON
RS-232 Interface. While the handler is establishing initial
communication, it is seeking a Ready response "R" from the tester. If anything other than an "
R
" or "
*
" is
received, the handler will pause and display the message, "
BAD RESPONSE
".
BAD TEST RESULT
The handler is using the E
XATRON
RS-232 Interface. After testing a device, the handler has received an
incorrect bin sort. You may press the
[Enter]
button to reissue a Start Test command and the handler will
try a second test. You may press the
[Clear]
button to abort the test and restart operation,
or
send the
handler an ASCII "
*
" or "
2A
" in HEX. This will automatically flush the DUT and begin a new test cycle. If
further details are required, please see Chapter 3, Interface Information.
BIN * S* C****
The shuttle is returning after properly placing a device in the output tube whose number is represented
above by an asterisk (
BIN *
). The handler had received a sort command represented above by an
asterisk (
S*
) from the tester/PC. The current device count in that output tube is represented above by four
asterisks (
C****
.)
Actual numbers will be displayed in place of the asterisks on your computer readout. For example:
The
message
BIN 1 S1 C0001
, indicates that the shuttle is returning after properly placing a device in output
tube 1, having received a sort 1 command and there is now 1 device in output tube 1. The counter is for
CURRENT TUBE FULL
count only.
The handler also keeps a
TOTAL
count. To obtain
TOTAL
counts, press the
[MANUAL]
button, then
press the
[TOTALS]
button for total counts from output bin keys
[1]
through
[8]
.
The tube full count rolls over at 9999 devices for output taping machine options.
BIN 5 S0 FLUSH
The handler is flushing one or more devices from the machine. This message is for the PC. It tells the
PC not to count the device(s) being flushed.
CHANGE RAM? NO
Exits routine without making any changes to the handler's RAM set-up parameters.
[Enter]
Exit with no changes.
[SET UP]
Toggle back to
CHANGE RAM ? YS
.
CHANGE RAM ? YES
Allows the operator to override the handler's set up parameters.
[Enter]
Allows changes to the RAM. Please refer to Chapter 4, Handler Set Up, for further
instructions.
[SET UP]
Toggle back to
CHANGE RAM ? NO
.