Exatron 3000B
www.exatron.com 5-9 Chapter 5 Diagnostics
;
;;
;
Tube Switch, 1 Through 8
__________________________________________________________________
This diagnostic tests port B6, located on PCB #5000-A89, PAL A89-8.
Please refer to drawing #6000-517 for sensor wiring details.
TO ENTER THIS CHECK:
Handler displays
DIAGNOSTICS
Press the
[1]
key.
Handler displays
SENSOR CHECK
Press the
[4]
key.
Handler displays
TUBE _ _ _ _ _ _ _ _
if all output tubes are removed from the handler. If numbers are
displayed, they indicate:
1 = Tube 1 Switch
2 = Tube 2 Switch
3 = Tube 3 Switch
4 = Tube 4 Switch
5 = Tube 5 Switch
6 = Tube 6 Switch
7 = Tube 7 Switch
8 = Tube 8 Switch
NOTE: The handler will display
TUBE CLEAR WORK
for two seconds whenever any of the output tubes
is removed. The message may flash when inserting a new tube. This is an indication that the "tube count
clearing" pulse is being received by the handler’s CPU. If this signal is not present, check pin 2,
PAL A89-10, on PCB #5000-A89. This pin should pulse whenever any tube is removed from the handler.
Pressing the
[Clear]
button will flash the terminal's displayed message to confirm the handler is still in this
specific diagnostic check.
Press the
[Enter]
button to EXIT this diagnostic check.