www.exatron.com 6-15 Chapter 6 Display Dictionary
OCTO/INDEX CHECK
Please refer to the Diagnostics/OCTO/INDEX CHECK section in Chapter 5 of this manual.
OUTPUT CLEAR ?
There has been a device jam at the handler's output. The handler now senses that the shuttle and the top
of the output tubes are clear of devices.
Open the handler's output door and verify that there are no jammed devices undetected by the handler's
sensors. If all is clear, close the output door and press the
[Enter]
button to resume operation.
OUTPUT TUBE JAM
The handler has sorted a device to an output tube whose LED is now flashing. The device is jammed at
the top of the tube.
Either clear the jam manually, (pressing the red solenoid override button to the right of the output tube
holder should do so) or remove the output tube and insert an empty tube.
PART DROPPED ?
The plunger vacuum may need to be strengthened since the plunger has dropped the device under test in
route to the test site.
Press the
[Clear]
button to flush the device into output tube 5.
Using a small screwdriver, slightly turn clockwise the yellow pot on the right side of the plunger vacuum
generator.
When the handler resumes operation, the small red light next to the yellow pot will light up if the plunger
vacuum is strong enough to properly hold onto the device under test.
PART IN PI SITE
This message applies to Type 2 and Type 3 Changeover Kits only. The fiber optic sensor in the P I
contact assembly has detected that a device has been left behind in the contacts by the pick-up head.
Manually remove the device. You may wish to undock the programmer from the handler to do so. Then,
check the test site vacuum sensor functionality by running "Test Site Sensors, 1 Through 8 Diagnostic" as
described in Chapter 5 of this manual.
PART IN SHUTTLE
The handler has detected a device in the shuttle which should not be there. The handler will automatically
flush the device to output tube 5.
PART IN T SITE
The handler has detected a device in the test site which should not be there. Press the
[Clear]
button to
flush the device into output tube 5.
PASS PROG
PASS VERIFY
The handler is using the PROGRAM RS-232 interface. The device under test has passed the
programmer's program/verify cycles. The current sum check will be displayed at the end of the above
messages.