www.exatron.com 5-6 Chapter 5 Diagnostics
;
;;
;
Test Site Sensors, 0 Through 7
________________________________________________________________
This diagnostic tests port 08, located on PCB #5000-398, PAL 398-72, U1, U2. Please refer to the Test
Head Wiring Diagram, in your Changeover Kit Supplement for sensor wiring details.
TO ENTER THIS CHECK:
Handler displays
DIAGNOSTICS
Press the
[1]
key.
Handler displays
SENSOR CHECK
Press the
[1]
key again.
Handler displays
SENSOR 0 67
for example, if the handler is empty and all sensors are working.
(Please refer to your Change-over Kit Supplement for specific sensor information and fill in the blanks
provided below.) The numbers shown in the display indicate:
0 Sensor #0 ________________
1 Sensor #1 ________________
2
Sensor #2 ________________
3
Sensor #3 ________________
4
Sensor #4 ________________
5
Sensor #5 ________________
6
Sensor #6 ________________
7 Sensor #7 ________________
Pressing the
[Clear]
button will flash the terminal's displayed message to confirm the handler is still in this
specific diagnostic check.
Press the
[Enter]
button to EXIT this diagnostic check.
If you suspect that you have a bad sensor/LED pair, use the E
XATRON
LED checker part #3000-052 (small
blue plastic box.) This diagnostic tool is equipped with a transistor (blue/white wires) and an IR LED
(orange/green wires.)
Direct the transistor at any of the handler's IR LEDs. The LED on the checker should turn off. Direct the
checker's LED at any of the handler's transistors. The corresponding number should turn OFF on the
handler's display.