Extron electronic 3000B Switch User Manual


 
Exatron 3000B
www.exatron.com 6-13 Chapter 6 Display Dictionary
MOTOR A BUSY ?
(Shuttle)
MOTOR B BUSY ?
(Z Motor Test Site Plunger)
MOTOR C BUSY ?
(Octoloader)
MOTOR D BUSY ?
(Optional - May Not Be Used)
The handler is checking to see if the stepper motor is moving as needed for each of the above handler
components. If this message does not go away within a few seconds, the handler will not operate. To
resolve this condition:
1. Turn the handler power OFF then ON and try again.
2. Check whether the motor setting address in RAM is correct: RAM ADDRESS 0091.
3. Verify that the motor control cable is plugged in and is not broken.
4. Replace the Motor Driver Module and/or the Motor Control PCB.
5. Contact E
XATRON
for assistance and replacement parts.
MOTOR A ERROR ?
(Shuttle)
MOTOR B ERROR ?
(Z Motor Test Site Plunger)
MOTOR C ERROR ?
(Octoloader)
MOTOR D ERROR ?
(Optional - May Not Be Used)
The handler is checking to see if the stepper motor is powering up properly as required for each of the
above handler components.
If this message does not go away within a few seconds, the handler will not operate. To resolve this
condition:
1. Turn the handler power OFF then ON and try again.
2. Check whether the motor setting address in RAM is correct: RAM ADDRESS 0091.
3. Verify that the motor control cable is plugged in and is not broken.
4. Replace the Motor Driver Module and/or the Motor Control PCB.
5. Contact E
XATRON
for assistance and replacement parts.
MOTOR OFF
The stepper motor being checked is now off.
MOVE MOTOR DOWN
The plunger motor has moved downward.
MOVE MOTOR UP
The plunger motor has moved upward.