Agilent Technologies 87130A Switch User Manual


 
Agilent 87130A Operating and Service Manual 4-65
Remote Operation
:VERify
ROUTE:VERIFY:ON ATTEN_14;
This command causes the set of switches defined by the PATH with the
pathname
ATTEN_14 to be added to (ON) or removed from (OFF) the "sense
list. All switches referenced in either channel list in the path are added (ON)
or removed (OFF).
ROUTE:VERIFY:OFF:ALL;
This turns sensing off for all channels.
Example Query ROUTE:VERIFY:ON? (@101,103,105);
This command returns a list of 1s and 0s, separated by commas, depending
on whether sensing is on or off for the indicated channels. For example, if
sensing is on for channels 101 and 105 and off for 103, the switch driver
responds with:
1,0,1
ROUTE:VERIFY:OFF? (@101,103,105);
This command returns a list of 1s and 0s, separated by commas, depending
on whether sensing is off or on for the indicated channels. For example, if it
is on for channels 101 and 105 and off for 103, the switch driver responds
with:
0,1,0
The query form is not available when using a path name, due to the potential
for confusion between the first group and the second in the path.