Agilent Technologies 87130A Switch User Manual


 
Agilent 87130A Operating and Service Manual 4-45
Remote Operation
:ERRor?
1004,EEROM data invalid
Corruption has been detected in the EEROM; it needs to be replaced, or, if
brand new, it needs to have
MEM:SAVE executed once.
1006,Channel timeout <17 hex digits>
This error is generated when a relay apparently fails to switch, based on the
detected state of the sense lines. The 17 hex digits indicate which card and
which of the 32 channels experienced failures just as for error 1001. Each
channel has two associated bits; the leftmost indicates that the error was
detected on the CLOSED side of the relay; the rightmost that the error was
detected on the OPEN side. A failure is indicated with a 1.
A Channel timeout without an accompanying Sense error
indicates that the sense lines were in a valid state, but it was the
wrong state.
A Channel timeout with a Sense error indicates that both sense
lines appeared to be at 0 V.
A Sense error without a Channel timeout means both sense lines
appeared to be at 24 V.
1007,Label too long
A path or group label was more than 32 characters long.
1008,Nonexistent group
An attempt was made to use a group name which was not previously set up
with a GROUP:NAME command.
1009,Group already exists
An attempt was made to name a group with a name already in use by another
group.
1010,Nonexistent path
An attempt was made to use a path name which was not previously set up
with a
PATH:DEF command.
4000,Divide by Zero
4001,Float Overflow
4002,Float Underflow
4003,Logarithm Error
4004,Integer Overflow
4005,Square Root Error
Query Errors
These set the Query Err bit in ESR.
-400,Query error
-410.Query INTERRUPTED
-420,Query UNTERMINATD
-430,Query DEADLOCKED