Introducing the 87130A Attenuator/Switch Driver
Before Installing the Attenuator/Switch Driver
Reducing ESD
Damage
To help reduce the amount of ESD damage that occurs during installation,
testing, or servicing instruments use the following guidelines:
Be sure that all instruments are properly earth-grounded to prevent
buildup of static charge.
Personnel should be grounded with a resistor-isolated wrist strap before
touching the center pin of any connector and before removing any
assembly from the instrument.
Before connecting any coaxial cable to an instrument connector for the
first time each day, momentarily ground the center and outer conductors
of the cable.
Handle all PC board assemblies and electronic components only at
static-safe work stations.
Store or transport PC board assemblies and electronic components in
static-shielding containers.
PC board assembly edge-connector contacts may be cleaned by using a
lintfree cloth with a solution of 80% electronics-grade isopropyl alcohol
and 20% deionized water. This procedure should be performed at a
static-safe work station.
Table 1-4 Static-Safe ESD Accessories
Part Number Description
9300-0797 Set includes:
3M static control mat 0.6 m x 1.2 m (2 ft x 4 ft) and 4.6 m (15 ft)
ground wire. (The wrist-strap and wrist-strap cord are not included.
They must be ordered separately.)
9300-0865 Ground wire, 4.6 m (15 ft)
9300-0980 Wrist-strap cord 1.5 m (5 ft)
9300-1367 Wrist-strap, color black, stainless steel, without cord, has four
adjustable links and a 7 mm post-type connection.
9300-1308 ESD heel-strap (reusable 6 to 12 months)
Order the above by calling an Agilent Sales and Service Office.