4-52 Agilent 87130A Operating and Service Manual
Remote Operation
:OPEN
:OPEN
Syntax
ROUTe:OPEN[?]
ROUTe:PFail:OPEN[?]
Description
Each channel has a CLOSE or OPEN position. On Agilent relays, the OPEN
path is the path between the input terminal labeled 1 on the relay and the
input terminal labeled C. It is recognized that CLOSE
and OPEN are
arbitrary for this type of switch; they are, however, in keeping with the SCPI
language specification.
Example Command ROUTE:OPEN (@101,2(0:5),3(1,3,5),406:410);
During any switching operation (OPEN or CLOSE) the settling bit in the
OPER status register is set (1). It is cleared (0) when the operation
completes.
Sending the above command causes channel 1 on card 1, channels 0 through
5 on card 2, channels 1, 3 and 5 on card 3, and channels 6 through 10 on
card 4 to be opened. Channels must be part of the drive list
(DRIVE:ON) to be
opened or they will be ignored.
Example Query ROUTE:OPEN? (@101,103,105);
Sending the query (optional question mark) causes the channels in the
channel list parameter to be checked for opened channels. The readback is a
list of 1s and 0s separated by commas, one for each channel in the list. A 1
is sent if the specified channel is opened, otherwise a 0 is sent. Although this
command is not strictly necessary (
ROUTE:CLOSE? is sufficient), it is
included for completeness.
Devices for which sensing is on, read back the position in which they were
sensed the last time a switching operation took place. Devices with sensing
off simply read back the currently programmed state (which will be the state
they were set to at power up if they havent been switched since power up).
Devices with
DRIVE:OFF read back the last value to which they were set.
<path name>
<channel list>



<path name>
<channel list>


