Agilent Technologies 87130A Switch User Manual


 
4-44 Agilent 87130A Operating and Service Manual
Remote Operation
:ERRor?
-181,Invalid outside macro definition
-183,Invalid inside macro definition
Execution Errors
These set the Exec Err bit in ESR.
-200,Execution error
-222,Data out of range
-223,Too much data
-270,Macro error
-272,Macro execution error
-273,Illegal macro label
-276,Macro recursion error
-277,Macro redefinition not allowed
Device Specific Errors
These set the Cmd Err bit in ESR.
-310,System error
-310,System error;SCPI internal
-310,System error;no memory
-310,System error;formatter
777,Queue Overflow
1001,Sense error;<17 hex digits>
This error is generated when an apparently impossible state of the sense lines
from one or more relays is detected. The first of the 17 hex digits represents
the card number on which the failure was detected; the rest represent a 64-bit
binary number which indicates which of the 32-channels experienced
failures.
Each channel has two associated bits; the leftmost indicates that the error
was detected on the CLOSed side of the relay; the rightmost that the error
was detected on the OPEN side. A failure is indicated with a 1. The least
significant (rightmost) two bits are channel 0, the most significant (leftmost)
are channel 31. If more than one card experiences failures then more than
one error will be reported.
1002,Memory capacity exceeded
An attempt has been made to store more data than will fit in the 16 Kbyte
nonvolatile memory area.
1003,Timer unstable
Instability has been detected in the 6840 timer chip, so the operating system
cannot run reliably.