Agilent Technologies 87130A Switch User Manual


 
4-14 Agilent 87130A Operating and Service Manual
Remote Operation
*ESE (Event Status Enable)
*ESE (Event Status Enable)
Syntax
*ESE mask
*ESE?
Description
The *ESE command sets the Standard Event Status Enable Register bits. This
register contains a mask value for the bits to be enabled in the Standard
Event Status Register. A one in the Standard Event Status Enable Register
will enable the corresponding bit in the Standard Event Status Register; a
zero will disable the bit.
Refer to the table below for the information about the Standard Event Status
Enable Register bits, bit weights, and what each bit masks. Valid number
range is 0 to 255.
The
*ESE query returns the current contents of the register.
Example Command OUTPUT 709;*ESE 64
Example Query OUTPUT 709;*ESE?
ENTER 709; Event
PRINT Event
Table 4-2 Event Status Enable Register Bit Definitions
Bit Weight Enables
7 128 PON- Power ON
6 64 (not used)
5 32 CME- Command Error
4 16 EXE- Execution Error
3 8 DDE- Device Dependent Error
2 4 QYE- Query Error
1 2 (not used)
0 1 OPC- Operation Complete