Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Calibration parameters
Calibration
In the menu "Analyzer Module Calibration" you can
start the zero calibration or start the span calibration
of all measurement ranges of the same channel for an MLT analyzer or analyzer module.
Measure Status... Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas: 50.00 ppm
Span gas name: Methane
Start zero calibration procedure !
Expected zero gas: 0.00 ppm
Valves...
Valve position: Samplegas
Time & Date: 12:50:55 June 16, 1998
Measure Status... Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Set Gas Valves --
Flow sample gas !
Flow zero gas !
Valve position: Samplegas
Operation status: Ready
Flow span gas !
Close all valves !
Flow test gas !
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Calibration