Main Menu — System configuration and diagnostics
Diagnostic menus
Control module diagnostics
In the menu "Control Unit Diagnostics" you can find software error messages of the
control module. If such messages are existing, you can act in the following way:
1) Make a note of the error messages.
2) Reset the message:
Press the 88 -key or the -key in the line "Edit to reset".
Set up "Reset" with the -key or the -key and confirm it with the 88 -key:
The error messages will disappear, if their reason will not exist any longer.
If the error messages will keep displayed:
3) System reset:
Change to the menu "System Reset" with the F5 key (Reset...).
Press the 88 -key or the -key in the line "System reset !" to reboot (see 6.5 p. 6-11).
4) Check the error messages:
Go again to the menu "Control Unit Diagnostics".
If the error messages will still exist, contact your customer service, please.
If the error messages will be disappeared, you should set up the report status again:
Set up "Report" in the line "Edit to reset".
Measure Back... Reset...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Control Unit Diagnostics --
Software error code (1 = no error): 1
And:
And:
And:
And:
And:
And:
Edit to reset: Report
Last software error message:
And:
6.1 Diagnostic Menus
6.1.1 Control Module Diagnostics