Emerson I-0021 Computer Accessories User Manual


 
4 - 38
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Measure Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Calibration Deviations --
Deviation from zero: 0.34 ppm
Deviation from span: 0.82 ppm
Sum of zero deviations: 4.41 ppm
Sum of span deviations: 7.57 ppm
Display Status... Main... Channel Calib...
TAG
F1 F2 F3 F4 F5
Failures: No
Maintenance-Requests: No
Temperature: 20.0 C
Operation: Ready
Range: 1
37.50 ppm CH4
0.00 50.00
0.0 100.0
Change to the single
component display of
the current channel
Press
F1
Options:
With the F3 key you can
change to further available
channels to check their
"Calibration Deviations".
With the F4 key you can
go back to the menu
"Calibration Procedure
Status".
Back to the single
component display