NGA-2000 System Calibration
Supplement - 16 NGA 2000 90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Here we first choose the specific analyzer module and then we select at the parameter
"Gastest of specific module" the required gas.
Each of the following types is possible:
SAMPLE-Gas (test mode deactivated)
Zero-Gas
All valves closed
SpanGas-1 (span gas for range 1)
SpanGas-2 (span gas for range 2)
SpanGas-3 (span gas for range 3)
SpanGas-4 (span gas for range 4)
With parameter "Timeout for Gastest" we can determine a time after which the activated
gastest switches automatically back onto SAMPLE-Gas.
Is this parameter set to "0" the automatic back switching is not done automatically and user
must break this mode with a "Cancel Calibration" Command.