Emerson I-0021 Computer Accessories User Manual


 
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 95
Select the tag of the module with the 88 -key or the -key:
The display will go back to the menu "DIO-Module Inputs" automatically.
The tag of the module selected will appear in the line "Signal comes from".
3) Select any function:
Change to the line "Choose function..." with the -key.
Press the 88 -key or the -key to change to the submenu "Functions":
Go to the line you want with the -key or the -key. With the F5 key you can go to
further menu pages containing also available signals:
SYS:Program-Cal
SYS:Cancel-Cal
SYS:CAL-Test-Mode
SYS:AM-Zero-Gas
SYS:AM-Span-Gas1, 2, 3, 4
External function control
AM: Hold Outputs
AM-closed-valves
Measure Back...
>>><<<
MLT25/CH2/R2
F1 F2 F3 F4 F5
95.00 ppm
-- Analyzer Modules --
Control Module
MLT25/CH1
MLT25/CH2
MLT25/CH3
MLT25/CH4
MLT25/CH5
Measure
>>><<<
Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Functions --
AM:Zero-Cal
AM:Span-Cal
AM:Range-1
AM:Range-2
AM:Range-3
SYS:Zero-Cal
SYS:Zero/Span-Cal
AM:Range-4
5.2 I/O Module Controls
5.2.2 DIO Module(s) – Input configuration