90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 97
• Go to the line you want with the ↑↑ -key or the ↓↓ -key. If there are more than eight
source modules available, you have to press the F5 key to go to another menu page.
• Select the tag of the module with the 88 -key or the →→ -key:
The display will go back to the menu "DIO-Module Outputs" automatically.
The tag of the module selected will appear in the line "Signal comes from".
3) Select any signal:
• Change to the line "Choose signal..." with the ↓↓ -key.
• Press the 88 -key or the →→ -key to change to the submenu "Signals":
• Go to the line you want with the ↓↓ -key or the ↑↑ -key. With the F5 key you can go to
further menu pages containing also available signals:
„ Range overflow
„ Range underflow
„ Flow too low
„ Flow too high
„ SYS:Valve-1, -2, ... , -32
„ and further signals, if need be:
All signals of the control module and of the analyzer modules joint to the platform or
MLT analyzer (e.g. CLD, FID, MLT AM) are available in the menu "Signals".
• Select the signal with the 88 -key or the →→ -key:
The display will go back to the menu "DIO-Module Outputs" automatically.
The signal selected will appear in the line "Signal name".
Measure
>>><<<
Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Signals --
Function control
Maintenance request
Failure
Cal. in progress
Zero in progress
Zero failed
Span failed
Span in progress
5.2 I/O Module Controls
5.2.2 DIO Module(s) – Output configuration