Emerson I-0021 Computer Accessories User Manual


 
4.8 Flow Span Gas
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 51
Display Status... Main... Channel Calib...
TAG
F1 F2 F3 F4 F5
Failures: No
Maintenance-Requests: No
Temperature: 20.0 C
Operation: Ready
0.0 100.0
Range: 1
37.50 ppm CH4
0.00 50.00
Measure Status... Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas: 50.00 ppm
Span gas name: Methane
Start zero calibration procedure !
Expected zero gas: 0.00 ppm
Valves...
Valve position: Samplegas
Time & Date: 12:50:55 June 16, 1998
Measure Status... Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Set Gas Valves --
Flow sample gas !
Flow zero gas !
Valve position: Samplegas
Operation status: Ready
Flow span gas !
Close all valves !
Flow test gas !
Change to the
menu "Analyzer
Module Calibration"
Press
F5
Caution:
Before starting span gas
flow, realize that spangas
is available !
(See also section 5.1.1,
p. 5-5... !)
Change to the
menu "Set Gas
Valves"
Press
F5
Change to the line
"Flow span gas !"
Press once to
get the line "Flow span
gas !" white on black.