Emerson I-0021 Computer Accessories User Manual


 
NGA-2000 System Calibration
Supplement - 26 NGA 2000 90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.2.1 Filling sequence buffer
The contents of the sequence buffer depends on
calibration types and belonging module type
coherence of modules and system valves
An evaluation procedure has the choice of listed action types.
Calibration types module type action types action data
NOP ALL-AMs USER-STEP Nr
ZERO-CAL single AM SWITCH_VALVE mask
SPAN-CAL Evaluation- PURGEWAIT time
ZERO/SPAN-CAL procedure ZERO AM
SPAN1-CAL SPAN AM range
SPAN2-CAL CALWAIT AM
SPAN3-CAL
SPAN4-CAL
END-OF-PGRM
One calibration type has following sequence frame:
1. SWITCH_VALVE
2. do different PURGEWAIT, ZERO, or SPAN which are possible with this valve adjustment
(order is determined by shortest purge time)
3. do different CALWAIT (wait for finishing a started calibration)
The evaluation procedure optimizes the order of actions by time.
The only condition is in case of ZERO/SPAN-CAL, where an AM has to do first its "ZERO"
before any of its "SPAN"s can be done.
The PURGEWAIT-delay-measurement is started with the last SWITCH_VALVE-action.
We can see, that the SYSCAL-types "ZERO_ALL" and "ZEROSPAN_ALL" are special cases
of a user program.
ZERO_ALL: ZERO-CAL of ALL-AMs
ZEROSPAN_ALL: ZERO/SPAN-CAL of ALL-AMs