Emerson I-0021 Computer Accessories User Manual


 
5 - 52
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Acknowledgement of status reports
In the menu "Acknowledgements" you can
acknowledge and clear all events of the MLT analyzer or analyzer module
starting the function "Clear events !" or
acknowledge and clear certain events with the functions
"Acknowledge and clear failures / maintenance requests / function controls !"
allow the acknowledgement and clear functions for the corresponding lines in the menu
"Status Details" (see section 4.1.1 p. 4-19 to 30)
set hours of operation to zero in the corresponding function line
set up hours for maintenance request in the corresponding line of variables.
Notes:
In the line "Hours of operation for maintenance request" you will set up, how many
hours the MLT analyzer or analyzer module may run until the maintenance will be
required. Options: 1 to 30000 hours.
The last line of the menu displays the total number of hours of operation since the
startup of an MLT analyzer or analyzer module resp. since the last reset of hours of
operation.
Via the F2 key you will reach the menu "Analyzer Channel Status". In that menu and its
submenus you will find the status report of the current channel for
Measure Status... Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Acknowledgement of Status Reports --
Clear events !
Acknowledge and clear failures !
Acknowledge and clear maintenance requests !
Acknowledge and clear function controls !
Set hours of operation to 0 !
Hours of operation: 164
Hours of operation for maintenance request: 720
Acknowledgement allowed in status menu: Yes
5.1 Analyzer Module Setup
5.1.9 Acknowledgement of Status Reports