Emerson I-0021 Computer Accessories User Manual


 
NGA-2000 System Calibration
Supplement - 32 NGA 2000 90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.4 Holding analog outputs of the SIO and avoid limit violation alarms
Any AM has LON-variable:
PROCESS
This variable is an input-variable. Its meaning is to tell the AM that another instance is not
allowing the measurement sample to flow valid.
In the AM are some states depending from PROCESS.
"valid measurement" only if PROCESS = "valid sample gas".
holding analog outputs if PROCESS = "no valid sample gas"
watching limit violations is switched off if PROCESS = "no valid sample gas"
It's now the task of the system calibration to handle the variable PROCESS for any involved
AM. This is done in the following way:
Any switching of a system valve looks if in the switched valves is a sample valve of any AM.
If yes, then it sets PROCESS like follows.
sample gas
valve
time delay PROCESS-
variable
switched off - sample not valid
switched on purge time of sample valve sample valid
It's very important to notice following:
SYSCAL only watches state of appropriate sample gas valve for setting of PROCESS-
variable.
Any additional valve, serial to sample valve, which can switch off flow of sample gas can not
be registrated for logic of PROCESS-variable.