Emerson I-0021 Computer Accessories User Manual


 
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 85
Analog Output Setup:
Press the 88 -key or the -key in the line "Analog output setup..." to change to the
corresponding submenu:
Configuration of the analog output:
1) Select the analog output number:
You can set up all parameters in the menu "Analog Outputs" separately for each of
the available analog outputs. The number of outputs depends on the hardware
configuration of the local SIO module (minimum: 2, maximum: 8).
Press the 88 -key or the -key to select the output number and adjust the value you
want with the -key or the -key.
Confirm the number selected with the 88 -key.
2) Choose the analyzer module:
Change with the -key to the line "Choose signal source module..."
Change to the submenu "Analyzer Modules" with the 88 -key or the -key.
The tags of all analyzer modules combined with the platform or the MLT analyzer will
appear:
Measure More...Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Analog Outputs --
Output number: 1
Choose signal...
Signal value for 0% output: 0.00 ppm
Signal value for 100% output: 2500.00 ppm
Output current range: 0...20 mA
Hold output during calibration? No
Current signal value: 95.00
Source module: TAG
Choose signal source module...
Signal name: Concentration
Measure Back...
>>><<<
MLT25/CH2/R2
F1 F2 F3 F4 F5
95.00 ppm
-- Analyzer Modules --
MLT25/CH1
MLT25/CH2
MLT25/CH3
MLT25/CH4
MLT25/CH5
5.2 I/O Module Controls
5.2.1 SIO Module – Analog output setup