Emerson I-0021 Computer Accessories User Manual


 
NGA-2000 System Calibration
Supplement - 30 NGA 2000 90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.2.3.5 Wait for Finishing Calibration
Any AM has the LON-variables
CALSTAT
Whether the selected AM is still calibrating or not, is checked by reading the variable
CALSTAT.
This CALSTAT-check is done in time intervals. During the intervals time is given to other
tasks.
4.2.4 Finishing System Calibration
After system calibration was working through the sequence buffer or it was canceled there is
to do some finishing work.
Actualize some displayed LON-variables
Look for some still running calibrations in case of cancelling and cancel them, too.
Restore modified LON-variables of AMs.
Switch system valves to sample gas state for all AMs
Cause to set any AM’s LON-variable PROCESS = "valid sample gas" after purging time of
sample gas valve.