NGA-2000 System Calibration
Supplement - 22 NGA 2000 90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
For example:
Starting time for zero calibration: daily at 2:00 o´clock.
Starting time for zero/span calibration: weekly at 10:00 o´clock.
Starting time for program calibration: weekly at 16.00 o´clock
Entry in menu page „times for interval operation“ are at 4/13/99 at 8.00 o´clock
First possible calibration: 4/14/99 at 2 o´clock (then all 24 hours automatically)
We can display the calculated ACT-times in menu page „next calibration events“...
But these times will only appear when the appropriate time controlled calibration type is
enabled.
Menu1Line
Menu2Line
Menu3Line
System configuration and diagnostics...
System Calibration...
Times for interval operation...
Next calibration events...
Next Calibration Events
Zero-Cal: 2:00 14 Apr 1999
Zero/Span-Cal: 10:00 13 Apr 1999
Program-Cal: 16:00 14 Apr 1999