90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 7
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Tolerances
↓↓
In the menu "Tolerances" you can set up the parameters of the stability controlled
calibration procedure (see p. 5-9/10) in one channel. Besides, you can disable the stability
control and the maximum calibration deviation (deviation control → deviation tolerance).
If you want to set up the parameters of the other channels of an MLT analyzer or analyzer
module, press the F3 key.
Set up parameters:
♦ Select any line of variables with the ↓↓ -key or the ↑↑ -key.
♦ Select the variable with the 88 -key or the →→ -key.
♦ Select any digit with the ←← -key or the →→ -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key
resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
♦ Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
Lines "Last zero calibration/Last span calibration":
The calibration status of the last calibration will be displayed in these two lines.
Measure Channel Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Tolerances --
Max. zero calibration deviation: 20.00 %
Stability tolerance - range 1: 100.00 %
Stability tolerance - range 2: 100.00 %
Stability tolerance - range 3: 100.00 %
Stability tolerance - range 4: 100.00 %
Last zero calibration: Success
Last span calibration: Success
Check deviation: Disabled
Max. span calibration deviation: 20.00 %
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Tolerances