Emerson I-0021 Computer Accessories User Manual


 
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 7
Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Calibration parameters
Tolerances
In the menu "Tolerances" you can set up the parameters of the stability controlled
calibration procedure (see p. 5-9/10) in one channel. Besides, you can disable the stability
control and the maximum calibration deviation (deviation control deviation tolerance).
If you want to set up the parameters of the other channels of an MLT analyzer or analyzer
module, press the F3 key.
Set up parameters:
Select any line of variables with the -key or the -key.
Select the variable with the 88 -key or the -key.
Select any digit with the -key or the -key and adjust a new value with
the -key or the -key
resp. select the whole parameter with the -key or the -key.
Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
Lines "Last zero calibration/Last span calibration":
The calibration status of the last calibration will be displayed in these two lines.
Measure Channel Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Tolerances --
Max. zero calibration deviation: 20.00 %
Stability tolerance - range 1: 100.00 %
Stability tolerance - range 2: 100.00 %
Stability tolerance - range 3: 100.00 %
Stability tolerance - range 4: 100.00 %
Last zero calibration: Success
Last span calibration: Success
Check deviation: Disabled
Max. span calibration deviation: 20.00 %
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Tolerances