Emerson I-0021 Computer Accessories User Manual


 
5 - 4
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
From the menu "Analyzer Module Setup and Controls" you can change to all menus and
submenus of the expert configuration level via the corresponding menu lines.
Changing to the menus and submenus:
Select the menu line you want with the -key or the -key.
If necessary: Change to a further menu page with the F5 key (More...)
Change to the corresponding menu/submenu with the 88 -key or the -key.
Menu "Load Factory Configuration":
If you press the F2 key (ManData) in the menu "Analyzer Module Setup and Controls (1/3)",
you will change to the following menu:
With the function of this menu you can delete the RAM data and load the factory settings
from the Flash-EPROM.
Caution:
This function is irreversible after starting (and confirmation) ! Your RAM settings
will be deleted !
Note:
The same function is available in the menu "Load/Save Analyzer Module Configuration" !
(see section 5.1.17 p. 5-64)
Starting the function:
Press the 88 -key or the -key or
cancel and go back to the last menu page with the F4 key (Back...) or the -key.
If asked: Confirm with the F2 key (Yes) or
cancel and go back to the last menu page with the F4 key (Back...) or the -key.
Further Ramifications via softkeys:
F1
Change to the single component display of the current channel !
+
Chapter 4 !
F3
Change to the other available channels
of an MLT analyzer or analyzer module !
+
See channel tag !
Measure Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Load Factory Configuration --
- BE CAREFUL with this function -
Replace current configuration with factory settings !