Emerson I-0021 Computer Accessories User Manual


 
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 17
Menu "Calibration Procedure Status":
Press the 88 -key or the -key in the line "Calibration procedure status..." to change to
the corresponding submenu:
In the menu "Calibration Procedure Status" you will find the results of the last calibrations.
This menu will appear automatically after starting the zero or span gas calibration. Then it
will display the status of the running calibration procedure (see sections 4.5/4.6 p. 4-39 to
4-46). If a calibration is still running, you can cancel it at any time with the F2 key.
With the F5 key (More...) you will change to a submenu, where you will find the calibration
deviations of the current channel. The deviations of the other channels of an MLT analyzer
or analyzer module you will find there by changing the channel with the F3 key.
Measure Cancel ! Back... More...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Calibration Procedure Status --
Procedure status: Ready
Current/expected gas flow: Samplegas
Concentration in span gas units: 37.50 ppm
Last zero calibration: Success
Last span calibration: Success
Last span calibration was: Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges: 1+2+3+4
Remaining procedure time: 0 s
Last zero calibration was: Fri 05-29-1998 13:32:06
---------------------- Results ----------------------
Measure Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Calibration Deviations --
Deviation from zero: 0.34 ppm
Deviation from span: 0.82 ppm
Sum of zero deviations: 4.41 ppm
Sum of span deviations: 7.57 ppm
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Calibration