Emerson I-0021 Computer Accessories User Manual


 
NGA-2000 System Calibration
Supplement - 14 NGA 2000 90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
3.4.1 Control via menu system
The possibility to control via menu is located in the menu page „Calibration/Test procedures“.
There can be started and stopped any kind of system calibration respectively test procedure.
SCTESTMOD
SCCALTYPE2
SCPROGSTEP
SCCALTIME1
SCCALTIME2
During a running SYSCAL there can be watched current information:
running calibration type
running program step of user defined program (other modes than user program show a
„0“!)
consumed calibration time
calibration time of last valid SYSCAL
System configuration and diagnostics...
System Calibration...
Calibration/Test procedures...
Calibration/Test Procedures
Start Zeroing all ranges!
Start Zeroing and Spanning all ranges !
Start calibration program !
Cancel calibration!
Calibration Procedures in Test-Mode: No
Test Procedures...
Calibration type: ZERO-CAL
Program step: 1
Calibration time: 16 s
Previous calibration time: 57 s
Result...