146 MicroMAX System User Manual Version 6.C
Acton (NTE with or without shutter)
1
Spacer Plate (removed)
Adapter (supplied with spectrometer)
Qty P/N Description
1. 3
2826-0127 Screw, 10-32 × 1/4, Button Head Allen Hex, Stainless Steel
Assembly Instructions
1. Make sure that the shipping cover has been rem
oved from the detector port on the
spectrometer.
2. Loosen the setscrews holding the Acton adap
ter in the spectrometer and remove the
adapter.
3. Rem
ove the spacer plate from the adapter by removing the three (3) socket head
screws.
4. Mount the Acton adapter to the face of the detector drum
housing (dashed outline in
illustration) with the three (3) 1/4" long button head screws.
5. Gently
insert the adapter into the spectrometer and fasten with the setscrews.
Note: Adapter parts are machined to provide a tight fit. It is necessary to rotate the detector
back and forth when inserting into the spectrometer adapter. Do not force the two parts of
the adapter together, as they can be permanently damaged by excessive force.