40 MicroMAX System User Manual Version 6.C
USB 2.0 Cable (6050-0494)
Turn the Controller power OFF (OFF = 0, ON = |) and the Computer power OFF before
connecting or disconnecting the Controller-Computer (TAXI) cable.
Caution
To Connect the USB 2.0 Cable:
1. Verify
that the Controller power is OFF.
2. Verify
that the Computer power is OFF.
3. Connect one end of the USB cable to the USB port on the host com
puter.
4. Connect the other end of the cable to the USB 2.0 port on the rear of the
Controller.
Connecting the Detector-Controller Cable
Turn the Controller power OFF (OFF = 0, ON = |) before connecting or disconnecting the
Detector-Controller cable.
Caution
To Connect the Detector-Controller Cable:
1. Verify
that the Controller power is OFF.
2. Connect m
ale end of the Detector-Controller cable to the “Detector” port on the back
of the Controller.
3. Move the slide latch over to lock the connector in place.
4. Connect the fem
ale end of the cable to the Camera.
5. Move the slide latch over to lock the connector in place.
Entering the Default Camera System Parameters into WinX
(WinView/32, WinSpec/32, or WinXTest/32)
Software changes implemented in WinX version 2.15.9.6 affected the way in which
default parameters were entered for camera systems. Therefore, two sets of instructions
are included. Follow the instructions appropriate to the software version that you
installed. Note that these instructions assume that you have performed the computer
interface installation.
WinX Versions 2.5.19.6 and later
1. Make sure the ST-133 is connected to the host com
puter and that it is turned on.
2. Run the WinX application. The Camera Detection wiz
ard will automatically run if
this is the first time you have installed a Princeton Instruments WinX application
(WinView/32, WinSpec/32, or WinXTest/32) and a supported camera. Otherwise, if
you installing a new camera type, click on the Launch Camera Detection
Wizard… button on the Controller/CCD tab page to start the wizard.
3. On the Welcome
dialog (Figure 11), leave the checkbox unselected and click on
Next
.