148 MicroMAX System User Manual Version 6.C
ISA HR 320 (NTE with or without shutter)
3
2
1
5
5
4
Remove spectrometer cover
for these screws.
Qty P/N Description
1. 1 2518-0106
Adapter-Female, HR320
2. 3
2826-0087 Screw, M5-10, Flat Head, Socket, Stainless Steel
3. 1 2518-0107
Adapter-Male, HR320
4. 3 2826-0127
Screw, 10-32 × 1/4, Button Head Allen Hex, Stainless Steel
5. 2
2826-0082 Set Screw, 10-32 × 1/4, Stainless Steel, Allen Hex, Nylon Tip
Assembly Instructions
1. Rem
ove the spectrometer cover.
2. Insert part 1 into the spectrom
eter (dashed line in illustration), fasten with the flathead
screws provided, and replace spectrometer cover.
3. Attach part 3 to the cam
era with the three (3) 1/4" long button head screws provided.
4. Gently
insert part 3 into part 1 and fasten with the setscrews.
3. Replace the spectrom
eter cover.
Note: Adapter parts are machined to provide a tight fit. It is necessary to rotate the camera
back and forth when inserting into the spectrometer adapter. Do not force the two parts of
the adapter together, as they can be permanently damaged by excessive force.