Xantrex Technology XMP 2600 Power Supply User Manual


 
Error codes Power module setup
APPENDIX
B
Error codes
The following is a list of error codes generated by the Power Modules and the main
controller of the XMP 2600.
Some of the main controller’s error codes pertain only to remote control.
Power
Modules
Errors Reported by the Power Module
1 Processor Watchdog was activated.
7 Severe OCP error - output did not shut down.
8 Severe OVP error - output did not shut down.
9 Test error - VOUT zero test failed.
10 Test error - half the full scale VOUT test failed.
11 Test error - OVP circuit did not trip.
12 Test error - OVP circuit did not reset.
13 Output did not shut down.
14 Sense lines voltage drop exceeded 4V.
15 Current Limit was active following a three seconds period of the output
being set to zero voltage during a Retry Foldback.
16 Test error - output current was not zero during the output test.
17 Internal Communication Time Out.
18 Internal Communications buffer overflow.
19 output voltage did not drop to zero during the reprogramming period when
the relays were operated.
20 Internal Communication Checksum error.
21 Illegal command in the Internal Communications.
22 Internal Communications Command Error.
Errors Detected by the Main Controller
78 Internal Communications Time Out.
251 Internal Communications response by the Power Module was in error.
Errors detected during Channels Scan at Power-On
240 Internal Communications Time Out - Power Module not installed.
XMP 2600 User’s Guide
rev. 1.0 93