Alarm Setup
Alarm Functions 5-15
Alarm Functions
The Nurse
Call Relay
Note—The nurse call relay is not available for the V24CT and V26CT.
The nurse call relay is a paging system which warns the nurse if an alarm
occurs on a monitor. It becomes active in one of four situations
depending on the way it has been configured:
Red only alarms
Red and Yellow alarms
Red and INOP alarms
Red and Yellow INOP alarms.
The nurse call relay is not active when the monitor’s alarm capability is
switched off.
Note—The delay time to the nurse call relay output is max. 1 sec.
When an alarm has been silenced using the key, and the
audible 2 or 3 minute alarm reminder is configured to be either On or Re-
alarm, the nurse call relay signals will behave as follows:
Alarm
Reminder
Reminder
Time
Nurse Call Relay Signal Behavior
On 2 or 3 mins The nurse call relay signals are repeated
for a duration of 6 seconds every 2 or 3
minutes unless the alarm condition is
eliminated
Re-alarm 2 or 3 mins The nurse call relay signals are activated
after 2 or 3 minutes. The alarm tone and
relay signals run continuously until they
are interrupted by pressing the
key.
Silence/Reset
Silence/Reset