Philips V26CT Computer Monitor User Manual


 
Controls and Connectors
Monitor Installation and Patient Safety 11-35
Monitor Installation
and Patient Safety
4. Philips monitoring network connector (SDN Interface); this is a
proprietary Philips system connector, used to input/output
information to a Philips monitoring network.
WarningWarning
Connecting the Philips monitoring network (SDN) cable when the
product is powered on is not supported. Error codes and Philips
monitoring network (SDN) interface lock-up may occur. Power
cycling the product will recover the product. No permanent
damage will result. To prevent unintentional disruption in
monitoring, be sure the SDN interface cable is properly secured at
both ends when connecting to the Philips monitoring network
(SDN).
5. Parameter Module Rack Connector; this connects the parameter
module rack to the monitor.
6. Defibrillator sync connector; sends an ECG signal out to a
defibrillator or other external device or receives a marker pulse
from a defibrillator or other external device. The marker pulse is
processed with the ECG signal and displayed on the monitor.
Note—To filter radiated emissions, a core shielding bead (Philips P/N
9170-1521) must be attached to the defibrillator synchronization cable
when a defibrillator is connected to the monitor. This filter has no impact
on the function of the monitor. To attach the core shielding bead:
a. Open the core shielding bead (filter).
b. Place the bead around the defib sync cable as close to the jack
as possible.
c. Close the bead over the cable and snap shut.
d. Plug the cable into the monitor's defib sync port. The bead
should be at the monitor end of the cable.