Philips V26CT Computer Monitor User Manual


 
Adjusting Neonatal Event Review Settings
9-20 Neonatal Event Review
Neonatal Event
Review
Bradycardia
Event Settings
Alarm Triggered
The Bradycardia event is triggered by a Red*** Brady alarm.
The Bradycardia alarm limit can be viewed but not adjusted in the Event
Setup task window. The alarm limit must be adjusted in Monitoring
Mode in the HR/PULSE or Alarm Limits task window and the trigger
time must be adjusted in Configuration Mode in the HR/PULSE task
window.
If the Bradycardia alarm related trigger threshold and trigger time items
are selected, the
key is inactive and the following
status message is displayed in the Event Setup task window:
“Brady items must be changed in the HR/PULSE Setup”
User Defined
For situations where the storage and documentation of an event should
be independent of all Bradycardia alarms, you can adjust the
Bradycardia event threshold and trigger time in Monitoring Mode in
the Event Setup task window.
Note—When the patient monitor is connected to an Arrhythmia System,
the Bradycardia alarms are generated by the latter and not the former.
The alarm limits can only be viewed and adjusted on the Arrhythmia. In
this situation, the Bradycardia threshold line is displayed in the Event
Review task window as a shaded area and the Brady trigger time and
threshold items are set to “From Arrhyth” in the Event Setup task
window.
Change Content