HP (Hewlett-Packard) E42 Switch User Manual


 
Model 5328A
Maintenance
Table 5-4. Performance Test (Continued)
.
Set the pulse generator for the following output:
a. (-) SLOPE TEST
Execute the following from the 9825A keyboard:
wrt 701,
“PF8G1S13A57+000*B57+000*R”
Counter should display 20 µS ±10 µs. Mark results on performance test record.
b.
(+) SLOPE TEST
.
Press RESET button on HP 5328A. Set Channels A and B slope switches to (-).
.
Execute the following from the 9825A keyboard.
wrt 701, “PF8G1S13A7+000*B7+000*R”
5-12